Are you in search of a reputable online distributor that can reliably source and ship top-notch military and defense parts, NSN parts, and aerospace parts? If so, we have you covered on Buy Mil-NSN Parts with our unrivaled inventory of offerings. Thank you for taking an interest in part number 4300M, a Daniel Consultants Inc item that is listed alongside the description “Analyzer Blood Gas” and the NSN 6630013648555. If you would like to learn more about this particular listing, you can always take the initial step of the procurement process via the submission of the quote request form on this page.
Note : We Will Not Share Your Information To Any Third Parties.
We organize our offerings in a variety of ways to make the search process simple for part number 4300M and other similar items. For instance, this Daniel Consultants Inc product is listed alongside the NSN 6630-01-364-8555, and it falls under Federal Supply Class (FSC) 6630 Chemical Analysis Instruments. Furthermore, through our rigorous vetting processes, careful part inspections, document verification practices, and more, we have established ourselves as a trusted distributor operating with AS9120B, ISO 9001:2015, and FAA AC 00-56B accreditation. With all of this in mind, begin the procurement process for part number 4300M from CAGE Code 0NTK1 on our website, and see why so many customers steadily depend on Buy Mil-NSN Parts for the NSN products they require.
MRC | Criteria | Characteristic |
---|---|---|
DPTH | Depth | 18.300 INCHES |
ELEC | Voltage In Volts | 120.0 AND 230.0 |
MRC | Decoded Requirement | Clear Text Reply |
CBBL | Features Provided | DIGITAL DISPLAY SCREEN AND INTEGRAL PRINTER AND SELF-CALIBRATION |
CXCY | Part Name Assigned By Controlling Agency | BLOOD GAS ANALYZER 120/230V 50/60 HZ AC: |
ACDC | Current Type | AC |
CYFV | Method Of Analysis | POLAROGRAPHIC AND POTENTIOMETRIC/ION SELECTIVE ELECTRODE |
ACZB | Frequency Rating | 50.0 HERTZ NOMINAL OR 60.0 HERTZ NOMINAL |
CYFY | Analysis Test/Measurement Range | PO2 0 TO 600 MMHG; PH 6.8 TO 7.8; PCO2 8 TO 99 MMHG; K 0.1 TO 9.9 MMOL/L; NA 100 TO 200 MMOL/L; CA 0.25 TO 4.25 MMOL/L; HCT 10 TO 75 PCT; HC03 3.0 60 MMOL/L TCO2 3.0 TO 60 MMOL/L; BASE EXCESS -30 TO +30 MMOL/L; O2 SAT 30 TO 90 PCT. |
CYHM | Test Sample/Specimen Type | WHOLE BLOOD |
ABRN | Height | 8.400 INCHES |
WDTH | Width | 9.100 INCHES |
AQXY | Test Type For Which Designed | BLOOD GASES AND ELECTROLYTES,HC03,HCT,BASE EXCESS AND 02 SAT |
CYFZ | Test Sample/Specimen Size | 0.5 ML |
FEAT | Special Features | UTILIZES A DISPOSABLE CARTRIDGE/TEST PACK THAT CONTAINS ALL ELECTRODES, CALIBRATING SOLUTIONS AND GASES PLUS A WASTE BAG FOR PROCESSED BLOOD, CALIBRATING AND QC SOLUTIONS. THE CARTRIDGE/TEST PACK SHALL ELIMINATE THE NEED FOR GAS TANKS EXTERNAL SOLUTIONS, AND ELECTRODE MAINTENANCE. SHALL BE PRE-PROGRAMED TO AUTOMATICALLY PERFORM TWO-POINT CALIBRATIONS EVERY HOUR AND ONE-POINT CALIBRATIONS EVERY 90 SECONDS. CAPABLE OF AUTOMATIC ASPIRATION OF THE SAMPLE AND TO PROCESS, DISPLAY, AND PRINT THE RESULTS IN LESS THAN 2 MINUTES. SHALL BE SUITABLE FOR THE FIELD ENVIRONMENT. |
If You’re Ever Looking for NSN parts Or Have an Aircraft On Ground Situation, Don’t forget That We offer Competitive Pricing and Guaranteed On-Time Delivery.
Request for Quote